Products

Water Quality Sensors
DRH890
Digital ORP Sensor
A digital ORP sensor that measures ORP and temperature.
Product Overview
DRH890is a digital sensor that measures ORP and temperature using an electrochemical method.
Suitable for the electronics, chemical, printing, dyeing, food, and pharmaceutical industries.
Supports immersion and flow cell installation in surface water, pure water, and circulating water systems.
Specifications
| Measured Parameters | ORP, Temperature |
|---|---|
| Measurement Principle | Electrochemical |
| Measurement Range | ORP: -2,000~+2,000 mV / Temperature: 0.0~60.0 °C |
| Resolution | ORP: ±1 mV / Temperature: 0.1 °C |
| Accuracy | ORP: 0.1% of range or ±2 mV, whichever is greater / Temperature: ±0.2 °C |
| Operating Temperature | 0~60 °C |
| Operating Pressure | ≤4 bar |
| Ingress Protection | IP68 |
| Power Requirements | DC 12 V |
| Power Consumption | Approx. 0.2 W |
| Electrical Isolation | Internal isolation between power supply and communication circuits |
| Communication Protocol | RS485 MODBUS / Modbus RTU |
| Temperature Compensation | Not separately specified in the manual |
| Calibration Method | Zero, slope, and offset calibration |
| Calibration & Maintenance | Visual inspection: monthly / Calibration check: monthly (depending on operating conditions) / Seal maintenance: every two years |
| Installation Type | Immersion, flow cell |
| Dimensions | 37 mm diameter / Approx. 261 mm overall length |
| Weight | Approx. 500 g (excluding cable) |
| Materials | 316 stainless steel / POM |
| Cable | PUR (polyurethane) jacket / Standard 10 m / Custom lengths available |
A better approach to water quality management starts with Water Eyes.
Prompt, accurate responses to product inquiries and technical questions.